| Measurement method |
| | BP-D5 | BP-D5L |
| |
Maximum bubble pressure method |
| Measurement range |
Surface tension: 10 - 100 mN/m, Lifetime: 20 - 5000ms (Fixing mode: 20 - 1000ms) |
| Resolution |
Surface tension: 0.01mN/m, Lifetime: 0.1ms |
| Repeatability |
0.5mN/m (standard deviation based on manufacturer's standard) |
| Measurement temp |
+10 - 70ºC (by optional hot/cold water circulator 4VT) |
| Sample volume |
About 80mL (exclusive lab dish is used.) |
| Sample viscosity |
0.5 - 10mPa · s |
| Dimensions (main body) |
320(W) x 315(D) x 475(H) mm |
321(W)×223(D)×422(H)mm |
| Weight (main body) |
18.5 kg |
11 kg |
| Utilities |
AC100 - 240V, 50/60Hz, 50W, 80VA, Pressurized dry air 0.4-0.7Mpa |
| Operating environment |
Room temp. 15~30℃, Humidity 30~80%RH non condensing |